NCICT 4 User Manual

Installation, inputs, phantom selection, calculation workflow, batch processing, and result interpretation for NCICT 4.

NCICT 4

NCI Dosimetry System for Computed Tomography

Documented version: 4.20261014 Update: October 14, 2026 — Scientific Update

This manual describes version 4.20261014 for macOS Universal (Intel and Apple Silicon) and Windows x64.

NCICT 4 main window showing phantom selection, scan coverage, scanner inputs, and organ dose output


Introduction

The National Cancer Institute Dosimetry System for Computed Tomography (NCICT) is a reference-grade radiation dose estimation system developed by the National Cancer Institute (NCI) for estimating organ absorbed doses and effective dose from computed tomography (CT) examinations.

NCICT is based on precalculated organ dose conversion coefficients derived from Monte Carlo radiation transport simulations, multiple computational human phantom models, and a reference CT scanner framework. The system supports patient-specific dose calculations, population-based dose evaluation, and benchmarking. It is not intended for clinical decision support.

The October 14, 2026 scientific update adds scanner-specific radial CTDI correction, cephalic/breech fetal presentation support, an expanded nCTDIw and nCTDIc/nCTDIp library, Monte Carlo uncertainty reporting, and corrected muscle and blood-vessel dose coefficients.


Calculation workflow

StepDescription
1Define patient characteristics
2Enter scanner parameters
3Define scan coverage
4Review organ and effective dose output
5Optionally run multiple cases using Batch Calculation

Numeric fields accept either dot or comma decimal notation regardless of the operating-system regional setting.


1. Patient characteristics

NCICT includes three phantom libraries to support matching of patient anatomy and body dimensions:

  • ICRP reference pediatric and adult phantoms (n=12)
  • Body size-dependent pediatric and adult phantoms (n=362)
  • Pregnant women and fetal phantoms at 8, 10, 15, 20, 25, 30, 35, and 38 weeks of gestation

1.1 Age-based selection: ICRP reference phantoms

When detailed body size information is unavailable, users may select one of the ICRP-defined reference age and sex groups:

  • 0-year (newborn) male or female
  • 1, 5, 10, or 15 years (male or female)
  • Adult male or female (assumed ≥20 years)

When ICRP phantoms are selected, reference height and weight are automatically assigned.

The ICRP reference height and weight pairs used by NCICT are:

Reference phantomHeight (cm)Weight (kg)
Newborn female or male474
1-year female or male7610
5-year female or male11019
10-year female or male14032
15-year female16153
15-year male16656
Adult female16860
Adult male17873

If a patient's age falls between reference groups, users may perform external dose interpolation or select the nearest reference age group.

ICRP reference phantom selection

1.2 Height- and weight-based selection: size-dependent phantom library

When patient height and weight are available, users may specify body size directly within the pediatric or adult phantom libraries. This approach maps the patient's body size to the closest matching size-dependent phantom. Phantom height and weight can be adjusted using the up/down arrow keys.

Height and weight input fields

Alternatively, users can click the "Phantom Height Weight Map" button to access an interactive size map. Blue cells represent available height/weight combinations, and the selected cell is outlined; clicking a cell updates the selection in the main panel.

Phantom height and weight map

1.3 Fetus tab

When fetal organ doses are of interest, users can select the "Fetus" tab and specify gestational age (8–38 weeks). Fetal anatomy is shown in color, while maternal anatomy is grayed out in the phantom display. The GUI lists the corresponding mother's perimeter (cm) next to each gestational age to help select the most appropriate pregnancy phantom. The grouped Cephalic and Breech radio buttons select fetal presentation; Breech is available at 10, 15, 20, and 25 weeks. On this tab, the first column of the organ-dose output table is labeled Fetal Organ.

Fetus phantom selection and Fetal Organ dose output

1.4 Mother tab

When maternal organ doses are of interest, gestational age and the corresponding mother's perimeter (cm) can be selected under the "Mother" tab. Maternal anatomy is shown in color, while fetal anatomy is grayed out in the phantom display. On this tab, the first column of the organ-dose output table is labeled Maternal Organ.

Mother phantom selection and Maternal Organ dose output


2. Scanner parameters

Two scanner parameter scenarios are supported.

2.1 CTDIvol unavailable

When CTDIvol is not available, it can be reconstructed from the scanner model, tube potential (kVp), tube current (mA), rotation time (sec), pitch, and total collimation (mm). CTDIvol (mGy) is automatically calculated using a built-in normalized CTDIw library in mGy/100 mAs for 160 selectable scanner/filter profiles. This profile count is not a count of unique scanner models because one scanner can have multiple filter conditions. Additional conditions are being added to the library through CTDI surveys.

Scanner information panel with calculated CTDIvol

Center and peripheral CTDI controls

The Scanner Information panel displays normalized weighted CTDI (nCTDIw), center CTDI (nCTDIc), and peripheral CTDI (nCTDIp) in mGy/100 mAs, where nCTDIw = (nCTDIc + 2 × nCTDIp) / 3. When component values are available for the selected scanner/filter, kVp, and CTDI phantom, NCICT loads them automatically. A • before the scanner name indicates built-in c/p values, and the line below the controls shows the CTDI profile ID and availability.

Normalized weighted, center, and peripheral CTDI controls

For example, CTDI-07-002-02 • Built-in nCTDIc/p available means:

  • CTDI-MM-SSS-FF is the permanent profile-ID format. MM identifies the manufacturer, SSS the scanner, and FF the filter (00 means no filter).
  • In the shown example, 07-002-02 maps to Siemens, Edge Plus, Standard in the current registry.
  • The ID identifies the selected scanner/filter condition; it is not a dose value.
  • • Built-in nCTDIc/p available means that center and peripheral values are available for the selected kVp and CTDI phantom and are loaded automatically.
  • If the availability text is absent, the profile does not provide a matching built-in pair. NCICT uses the conventional CTDIvol-only result unless the user supplies a validated custom pair.

Users may edit nCTDIc or nCTDIp. NCICT keeps nCTDIw fixed and updates the other component automatically. Pressing Enter recalculates the doses, and custom values are shown in blue. Reset restores the built-in values or the reference defaults. Custom values should be used only when appropriate validated measurements are available.

2.2 CTDIvol available

Users may directly enter CTDIvol and select the CTDI phantom type (head CTDI phantom with 16 cm diameter or body CTDI phantom with 32 cm diameter). The entered value overrides the scanner-derived CTDIvol calculation. When an eligible CTDI profile or validated nCTDIc/nCTDIp pair is available, NCICT can still apply the scanner-specific organ-dose correction; otherwise it uses the conventional CTDIvol-only calculation.

Scanner information panel with CTDIvol input

2.3 Radial organ dose correction

For an eligible scanner profile or validated custom nCTDIc/nCTDIp pair, NCICT uses the relative center-to-periphery dose distribution together with the phantom- and organ-specific radial position to correct organ and effective doses. The correction preserves nCTDIw, while CTDIvol remains the absolute dose input; it does not replace or recalculate CTDIvol. If an eligible pair is not available, NCICT uses the conventional CTDIvol-only calculation.

The adult-body method was validated against scanner-specific Monte Carlo organ-dose coefficients from Jansen et al. (2022) using 35 adult-body scanner-configuration/tube-potential conditions from eight scanner models at 80, 100, 120, and 140 kV. Each condition was evaluated in both ICRP adult male and female phantoms, for 70 Monte Carlo datasets in total. Median interscanner organ-dose variation decreased from approximately 8% after conventional CTDIw normalization to approximately 2% after radial correction. The associated manuscript has been submitted for peer review. These findings apply to eligible 32-cm adult-body conditions and do not represent total patient-specific dose uncertainty.

2.4 Tube current modulation

A generic longitudinal tube current modulation (TCM) model is implemented for illustrative purposes. It is not vendor-specific. TCM strength ranges from 0 (no modulation) to 1 (maximum modulation). The modulated current profile is displayed next to the phantom panel.

  • DLP is calculated from CTDIvol and scan length
  • WED is the mean of positive slice WED values over the selected scan range; slices without phantom tissue (WED = 0) are excluded
  • SSDE is calculated using regression equations based on this mean WED

These values are displayed together with the scan start, scan end, scan length, and predefined protocol below the phantom display.

CT dose metric outputs


3. Scan coverage definition

Scan coverage can be defined in the Scan Range and Dose Metrics panel:

  • interactively (mouse drag)
  • numerically (Scan Start cm / Scan End cm)
  • using the up/down arrow keys in the text fields
  • using predefined protocols

After a user-defined scan range is entered, NCICT maintains the corresponding anatomical range when switching among the ICRP, Size, Fetus, and Mother tabs. Selecting a predefined protocol replaces the custom range with that protocol's range.

Phantom display with scan coverage box Scan coverage controls


4. Dose output

NCICT reports organ absorbed dose (mGy) and effective dose (mSv). Results are intended for comparative and population-level analyses, not individual risk estimation.

Detailed cardiac substructure doses are available only for NCI size-dependent phantoms. The pregnant phantoms do not contain separately tallied cardiac substructures: the GUI hides these rows, and Batch output leaves the corresponding cells blank. The reported Heart wall dose remains available and is not copied into the unavailable substructure fields.

The GUI shows doses to three decimal places and positive doses below 0.001 as <0.001. MC uncertainty is shown to two decimal places. Batch CSV and API JSON retain numeric dose precision for subsequent calculations; MC uncertainty is available only in the GUI.

Organ dose output table

4.1 Monte Carlo statistical uncertainty

The MC uncert % column reports the relative Monte Carlo counting uncertainty as one standard deviation. This is a statistical uncertainty from the dose-coefficient simulations. It does not include uncertainty from patient or phantom anatomy, scanner output, CTDI measurements, x-ray spectrum and transport models, cross sections, or other systematic sources.

NCICT propagates the available slice-level relative uncertainties through spectrum interpolation and sums independent slice variances in quadrature. Dose uncertainty for active marrow and shallow marrow is first propagated from the contributing skeletal-site and photon-energy fluence tallies using the corresponding dose-response coefficients and marrow-distribution fractions. Effective-dose uncertainty is propagated from the ICRP Publication 103 tissue-weighted dose components. MCNP covariance terms were not retained in the dose library, so component tallies and source slices are treated as independent in these calculations.

Dose coverage is the fraction of the displayed organ dose contributed by slices for which finite MC uncertainty data are available; it is not the fraction of slices. If D_i is a slice dose contribution and r_i its relative MC uncertainty, the partial-data quantities are:

dose coverage = sum(D_i with uncertainty) / sum(D_i from all slices)
reported relative uncertainty =
    sqrt(sum((D_i × r_i)^2 for slices with uncertainty)) / sum(D_i from all slices)

The displayed organ dose always includes all selected scan slices, including slices without uncertainty data. Only unavailable variance terms are omitted. The output uses a fixed 90% minimum dose-coverage quality threshold:

Every numeric MC uncertainty reported by NCICT is calculated from uncertainty data covering at least 90% of the displayed organ dose.

  • 9.42 means that at least 90% of the displayed dose has usable uncertainty data and the reported relative MC uncertainty is 9.42%. NCICT uses all available uncertainty data; it does not truncate a result at 90% coverage.
  • N/A means dose coverage is below 90%, or no usable MC uncertainty is available. NCICT continues to show the library dose result; N/A applies only to the MC uncertainty column.

When coverage is between 90% and 100%, unavailable variance terms are omitted, so the reported value may slightly underestimate the complete MC statistical uncertainty. The 90% rule is intended as a practical reporting threshold for communicating the statistical reliability of the dose estimate, not as a complete uncertainty budget.

The 90% threshold is fixed so the same calculation is reported consistently for all users and in repeated analyses; it is not a user-adjustable setting.

Cardiac substructure acronyms

AcronymCardiac Substructure
H LALeft Atrium
H RARight Atrium
H LVLeft Ventricle
H RVRight Ventricle
H LVMLeft Ventricular Myocardium
H LMCALeft Main Coronary Artery
H LADALeft Anterior Descending Artery
H LCALeft Circumflex Artery
H RCARight Coronary Artery
H AVAortic Valve
H MVMitral Valve
H PVPulmonary Valve
H TVTricuspid Valve

5. Batch calculation mode

Input parameters can be provided in CSV format. A template file (ncictBatchInput.csv) is included with the NCICT distribution and can be edited in Excel.

5.1 Workflow

  1. Make a copy of ncictBatchInput.csv, keep the header row, and enter one examination per row.
  2. In NCICT, click Batch Calculation and select the input CSV.
  3. In the Batch output dialog, choose the output location and filename. The suggested name is ncict_batch_output.csv.
  4. NCICT calculates all valid rows and reports the processed and skipped counts. The saved output contains the successfully processed rows.

Cardiac substructure columns remain in the fixed Batch output schema. For fetus and mother calculations, these cells are blank because the pregnant phantoms do not contain separately tallied cardiac substructures. A blank cell means unavailable, not zero; Heart wall dose is still reported separately.

If a required header is missing, calculation stops with an error message. If individual rows are skipped, the completion message shows the last row error. Common row errors include an unknown profile ID, a profile that does not match the submitted kVp or CTDI phantom, a partial nctdic/nctdip pair, or invalid numeric values.

5.2 Input file and columns

Batch input accepts comma- or semicolon-delimited CSV files. Semicolon-delimited CSV is recommended when decimal commas are used. In a comma-delimited file, a value containing a decimal comma must be enclosed in double quotes. Saved Batch CSV output always uses comma delimiters and dot decimals for consistent reuse across regional settings.

Recommended Batch CSV column order:

dose_target,age,sex,presentation,height,weight,wed,start,end,kvp,tcm_strength,head_body,ctdivol,ctdi_profile_id,nctdic,nctdip,protocol,custom_ma

The Batch parser reads column headers, so columns may be reordered as long as the required headers are present. Required headers are dose_target, sex, kvp, tcm_strength, head_body, and ctdivol, plus either protocol or both start and end. Each row must provide a protocol name or both boundaries. The age column is optional for patient-only files but is required when a row requests fetus or mother dose. The sex cell may be blank for fetus or mother dose. Additional empty columns may be placed after custom_ma to enter slice-specific custom mA values across Excel cells.

CategoryColumnUse
Case identifierpatientIDOptional case identifier; rows are numbered if omitted
Phantom selectiondose_targetRequired dose recipient: patient, fetus, or mother
Phantom selectionageAt least one required*: patient age in years; gestational age is always required for fetus or mother
Phantom selectionsexRequired for patient dose; f, m, female, and male are accepted; ignored for fetus or mother
Phantom selectionpresentationOptional cephalic or breech; defaults to cephalic and is ignored for patient dose
Phantom selectionheightAt least one required*: patient height in cm; ignored for fetus and mother
Phantom selectionweightAt least one required*: patient weight in kg; ignored for fetus and mother
Phantom selectionwedAt least one required*: scan-average water-equivalent diameter in cm; leave blank when unavailable; ignored when height or weight is available and for fetus and mother
Scan rangeprotocolOptional predefined protocol name with alias and minor-typo matching; overrides start/end
Scan rangestart, endRequired when protocol is blank: distance from the top of the head in cm, or anatomical landmark IDs
Scanner parameterskvpTube potential in kVp; must be greater than 0
Scanner parameterstcm_strength0 to 1; leave blank when providing a custom mA profile
Scanner parametershead_body1 = 16 cm head CTDI phantom; 2 = 32 cm body CTDI phantom
Scanner parametersctdivolRequired CTDIvol in mGy; must be greater than or equal to 0
Scanner parametersctdi_profile_idOptional exact profile ID for the submitted kVp and CTDI phantom
Scanner parametersnctdic, nctdipOptional center/peripheral CTDI in mGy/100 mAs; enter both as positive values
Scanner parameterscustom_maFirst custom mA value; required when tcm_strength is blank

* For patient dose, provide at least one of age, height, weight, or wed.

5.3 CTDI input modes

ctdivol remains the required absolute dose input in every mode.

ModeInput fieldsOutput radial_mode
CTDIvol onlyLeave ctdi_profile_id, nctdic, and nctdip blankctdivol_only
Built-in scanner profileEnter the exact ctdi_profile_id; leave nctdic and nctdip blankscanner_library
Explicit center/peripheral valuesEnter positive nctdic and nctdip together; the ID is optional provenanceexplicit_cp

Profile IDs can be read from the GUI status line or from ncictCtdiProfiles.csv. A profile ID must have data for the submitted kvp and head_body. Explicit nctdic and nctdip values use mGy/100 mAs and take priority when both an ID and a pair are present.

Profile components are linearly interpolated between available kVp values. A supplied ID must still be recognized when an explicit pair is provided. With no profile or pair, Batch applies no scanner-specific correction.

p_over_c is calculated internally from nctdip/nctdic; it is not an input column. The included template contains examples of all three modes.

5.4 Patient and phantom matching

Dose targetAvailable informationPhantom selection
PatientHeight and weightClosest size phantom
PatientHeight onlyEstimate weight from closest sex-specific ICRP reference
PatientWeight onlyEstimate height from closest sex-specific ICRP reference
PatientWED onlyPhantom with the closest WED over the scan range
PatientAge and sex onlyAge- and sex-specific reference body size
PatientAge unavailable, body size availableInfer the closest age/size group
PatientAge and body size unavailableInput error: provide age, height, weight, or WED
FetusGestational ageFetal phantom for the closest supported week
MotherGestational ageMaternal phantom for the closest supported week

Additional rules:

  • Patient size priority: height/weight → WED → age/sex reference.
  • For patient dose, at least one of age, height, weight, or wed is required.
  • If WED and height/weight are both provided, WED is ignored.
  • Gestational age is required for fetus and mother.
  • Sex, height, weight, and WED are ignored for fetus and mother.
  • If presentation is omitted, cephalic is used.

Patient Batch rows select NCI size-dependent phantoms. Reference body sizes are matching targets; they do not select the GUI's ICRP voxel phantoms. Patient age may be fractional, and 0 is valid for a newborn. Height, weight, and WED must be positive to count as available body-size data. For mother dose, the maternal library uses cephalic presentation, including when breech is submitted.

Patient age below 20 years selects the pediatric candidate group; age 20 years or older selects the adult group. If age is omitted, NCICT searches the same-sex pediatric and adult candidates using the available body-size information. If age and body-size information are both omitted, the row is skipped with an input error.

For WED matching, NCICT calculates each candidate phantom's mean WED over the submitted scan range and selects the closest value. WED is supported for landmark-based and custom scan ranges and is not converted to body weight.

Supported gestational ages are 8, 10, 15, 20, 25, 30, 35, and 38 weeks. Other positive gestational ages are matched to the nearest supported week; ties select the lower week. Breech fetal phantoms are available at 10, 15, 20, and 25 weeks. For other weeks, cephalic presentation is used.

5.5 Protocol names, scan range, and landmarks

Scan-range priority is protocol name → start/end. A nonempty protocol selects its predefined anatomical range and overrides both boundaries; the start and end cells may be blank. Without a protocol, both boundaries are required in cm or as landmark IDs.

Supported names match the GUI: Head, Neck, Chest, Abdomen, Pelvis, AP (abdomen-pelvis), and CAP (chest-abdomen-pelvis). Matching ignores case, spaces, common separators, and the words CT, scan, and protocol. Common aliases include brain, thorax, abdomen/pelvis, and abdominopelvic; minor spelling errors such as chset CT also match. Unknown or ambiguous names skip the row with an error instead of falling back to its start/end values.

WED-based selection uses the resolved protocol range for each candidate phantom. Output records the submitted name in protocol, the resolved name in matched_protocol, and the final start/end in cm. Both name columns are blank when only start/end were supplied.

Output protocolOutput matched_protocol
ChestChest
chset CTChest
abdomen/pelvisAP
CAPCAP
Keep protocol before custom_ma so cells following custom_ma contain only
custom mA samples.

Scan start and end locations can be specified either as distances from the top of the head in cm or as anatomical landmark IDs. If a landmark ID is used for one boundary, a landmark ID should also be used for the other boundary.

Anatomical landmark IDs

Common protocol ranges can be entered using the following landmark ID pairs:

ProtocolStart landmark IDEnd landmark ID
Head10011003
Neck10021005
Chest10041007
Abdomen10061008
Pelvis10081009
Abdomen-pelvis10061009
Chest-abdomen-pelvis10041009

Other valid landmark pairs, including 1001 to 1010, may be entered directly for customized scan ranges.

After landmark conversion, the scan range must satisfy 0 <= start < end <= 190.

5.6 Custom mA profiles

When editing the Batch CSV in Excel, custom mA values can be entered across multiple cells starting at the custom_ma column:

dose_targetagesexpresentationheightweightwedstartendkvptcm_strengthhead_bodyctdivolcustom_ma
patient10f25020120120100200300400500600700

Enter each custom mA value in its own Excel cell rather than as an array with brackets. NCICT reads the custom_ma cell and all numeric cells to its right as a single custom mA profile. The delimiter and decimal-separator rules in Section 5.2 also apply. Leave these cells blank when a numeric tcm_strength from 0 to 1 is used.

5.7 Examples and output

The included test template has 38 cases: 33 expected successes, one intentional missing-patient-information failure, and four protocol-matching failures (Whole body, Cardiac, Chest abdomen, and xyz). Running it should report 33 processed and five skipped rows. Remove the intentional failure rows when using the template for routine calculations.

Protocol names can replace start/end entirely:

dose_target,age,sex,protocol,kvp,tcm_strength,head_body,ctdivol
patient,40,m,Chest CT,120,0,2,10
mother,20wk,,abdomen/pelvis,120,0,2,10

These rows show CTDIvol-only, built-in profile, explicit c/p, fetus, and custom mA cases:

dose_target,age,sex,presentation,height,weight,wed,start,end,kvp,tcm_strength,head_body,ctdivol,ctdi_profile_id,nctdic,nctdip,custom_ma,,,,,,
patient,10,f,,,,,0,20,120,0,1,20,,,,
patient,20,m,,175,80,,0,20,120,0,2,20,CTDI-07-002-02,,,
patient,20,m,,175,80,,0,20,120,0,2,20,,3.84,6.98,
fetus,38wk,,,,,,1004,1009,120,0,2,20,,,,
patient,10,f,,,,25,0,20,120,,1,20,,,,100,200,300,400,500,600,700

Output CSV includes:

  • Patient ID and dose target
  • Age, sex, presentation, height, weight, WED, and scan start/end
  • Scanner parameters
  • Radial mode, profile ID, resolved nCTDIc/nCTDIp, p/c ratio, and CTDI library version
  • Matched phantom group and matched phantom size or fetal age
  • Submitted protocol and resolved matched_protocol, with final scan start/end in cm
  • Organ doses and effective dose

The output order is phantom_group, matched_height_cm, matched_weight_kg, matched_fetal_age_weeks, protocol, matched_protocol, followed by organ doses. Dose values retain calculated numeric precision; format these columns in Excel for the desired display.

When WED is not provided, its output CSV cell is left blank; the internal missing-value marker is never written to the user-facing file. WED is also blank when ignored because height/weight or a pregnant dose target was supplied. For a custom mA profile, the tcm_strength output cell is also left blank.